Fitting
1. Turn speech processor, FM receiver and transmitter off.
2. Plug the orange cable into the MicroLink CI S.
3. Plug the MicroMLxS into the MicroLink CI S. Leave the receiv-
er in the double green dot position.
4. Attach the MicroLink CI S to the external input socket on the
speech processor.
5. Turn on the speech processor, FM transmitter and MicroLink
CI S in that order to avoid transient sounds.
6. Select a program that allows for sensitivity adjustment.
7. Determine speech perception in quiet with the CI alone at
normal (12) sensitivity. Repeat test with the FM system. If
scores are poorer with the FM system, increase the gain of the
MicroLink CI S until comparable performance is achieved.
8. Determine speech perception with the FM system in noise.
Decrease mic sensitivity in two-unit increments until signifi-
cant FM benefit is demonstrated.
Troubleshooting
Listening check #1
The environmental and FM signal can be heard simultaneously
through the monitoring phones to confirm function of the head-
piece mic and the FM system.
Listening check #2
1. Plug the receiver and adapter into a test amplifier, e.g. Radio
Shack speaker, with the orange patch cable.
2. Insert a battery into the MicroLink CI S and turn it on.
3. Put the MicroMLxS receiver in the double green dot position.
4. Turn on the FM transmitter. While speaking into the transmit-
ter, listen for the clear signal with minimal background noise.
This check will not reflect the sound quality that the patient
will hear but will confirm function and transmission of the
FM system.
5. To check the FM system, remove the MicroMLxS and plug it
into a working hearing aid and audioshoe or a Phonak
MicroLink Headset checker. Turn on the transmitter and listen
for a clear signal.
Listening check #3
1. Follow your audiologist’s recommendations for conducting a
listening check of your speech processor.
NOTE: You may choose to disable the volume control. If both
the microphone sensitivity and the volume controls are
available, press the Select button to alternate between
them.
Required Parts:
1. MicroMLxS
2. MicroLink CI S
3. Orange cable
1
2
3
MicroLink CI S plugged into
the RadioShack speaker
Monitoring Earphones
Control buttons
Program
Segment meter
Microphone
sensitivity
Options button
Select button
Autosensitivity
button
On/Off button
Up
button
Down
button
Program
button
SPrint Cochlear
Headset Checker
(Phonak part number 052-3126
[110V] and 052-3122 [220V])
Page view 12